SM-FTN Needle move limit exceed (5.0) - WKB288265

Article number: 288265

SYMPTOMS

  • Needle move limit exceed (5.0)

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle  (SM-FTN)
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS

CAUSE

Bent or damaged sample needle

FIX or WORKAROUND

  1. Replace the sample needle.

ADDITIONAL INFORMATION

  1. Perform Needle Z-axis calibration.
  2. Perform Needle seal characterization.
  3. Perform the needle seal readiness test.