SM-FTN Needle move limit exceed (5.0) - WKB288265
Article number: 288265
SYMPTOMS
- Needle move limit exceed (5.0)
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
CAUSE
Bent or damaged sample needle.
FIX or WORKAROUND
- Replace sample needle
- Perform Needle Z-axis calibration
- Perform Needle seal characterization
- Perform needle seal readiness test
ADDITIONAL INFORMATION
id288265, UPARBSMFTN, UPARHSMFTN, UPARPFTN, UPARSMFTN, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO