ACQUITY PDA fails SystemsQT drift test - WKB28762

Article number: 28762

SYMPTOMS

  • Drift test fails
  • Energy is OK

ENVIRONMENT

  • ACQUITY UPLC PDA Detector

CAUSE

Contaminated flow cell

FIX or WORKAROUND

  1. Flush the flow cell with a compatible solvent such as 100% acetonitrile at 1.0 - 2.0 mL/min for about an hour.
  2. Repeat the Noise and Drift test.

ADDITIONAL INFORMATION

id28762, A-30PDA, EMP2QLIC, EMP3Q1ULIC, EMP3QLIC, UPAPDA, UPBINARY, UPC2PDA, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.