Sudden pressure drop with ACQUITY UPLC I-Class Binary Solvent Manager - WKB287413

Article number: 287413

SYMPTOMS

  • Loss of flow/pressure
  • Solvent lines are primed, both through ACQUITY Console and manually
  • Pressure restored when MeOH is used as mobile phase
  • Pressure drops when ACN is used as mobile phase

ENVIRONMENT

  • ACQUITY UPLC I-Class Binary Solvent Manager 
  • ACQUITY UPLC I-Class Binary Solvent Manager PLUS

CAUSE

Faulty primary check valve

FIX or WORKAROUND

Replace the primary check valve (part 700004139).