Unstable system pressure with ACQUITY Arc or Arc HPLC - WKB286127

Article number: 286127

SYMPTOMS

  • Pressure unstable and no stable delta can be reached
  • Pressure is lower than 1000 psi

ENVIRONMENT

  • ACQUITY Arc
  • ACQUITY Arc Bio
  • Arc HPLC 
  • QSM-R

CAUSE

Operating below system specification

FIX or WORKAROUND

Optimize the column selected for the method, or use a restrictor to increase system pressure to at least 1000 psi.

ADDITIONAL INFORMATION

See page 136 of the ACQUITY Arc, ACQUITY Arc Bio, and Arc HPLC Quaternary Solvent Manager-R Overview and Maintenance Guide for QSM specifications.