ACQUITY UPLC SM-FTN sample loop leak/void causes poor peak shape - WKB284909
Article number: 284909
SYMPTOMS
- Poor peak shape
- System leak tests and sample manager diagnostics pass
- Injection valve was recently replaced
- Volumes in the console are correct
- Pressure increases slightly with no sample loop installed
ENVIRONMENT
- ACQUITY UPLC SM-FTN
- I-Class
- H-Class
CAUSE
Sample loop has a small leak or void volume.
FIX or WORKAROUND
- Remove and reseat the sample loop in the injection valve.
- If poor peak shape persists after reseating, replace the sample loop.
ADDITIONAL INFORMATION
Part numbers for sample loops:
I-Class/H-Class
- Sample Loop, 50-µL (430002012)
- Sample Loop, 100-µL (430002011)
- Sample Loop, 250-µL (430002007)
H-Class Bio
- Sample Loop, 50-µL (430002591)
- Sample Loop, 100-µL (430002625)
- Sample Loop, 250-µL (430002630)
- Sample Loop, 1000-µL (430002585)