"Waiting for injection start" message with MassLynx and Autopurification - WKB2818

Article number: 2818

SYMPTOMS

  • MassLynx presents the following message in the bottom-right of the sample list window: "waiting for injection start"
  • The acquisition does not start
  • Error in Status Log in Collection Control "[1][---] [---] - Timed out waiting for raw data, check the data file"

ENVIRONMENT

  • MassLynx
  • Autopurification
  • 3767

CAUSE

The default software settings must be changed to accommodate Autopurification.

FIX or WORKAROUND

Amend the triggering settings in the inlet editor:

  1. Open the inlet editor.
  2. Click on Tools.
  3. Click on Instrument Configuration.
  4. Click on Events and Triggering.
  5. Set the pump to Trigger by Software (see screenshot below).

clipboard_ea0dd20c005e1432ebcf5afad2c1a62ef

Newer purification system with 3767 must be configured as follows:

clipboard_ef764c42e273d9b934cd7054affc77f59.png

ADDITIONAL INFORMATION