Error: Plunger Homing Over Pressure in Alliance 2695 Separations Module - WKB2816
Article number: 2816
SYMPTOMS
- Plunger homing over pressure (0) or (1) error on startup
- Power-off system, open vent valve, power-on system, error did not occur
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2795 Separations Module
CAUSE
Clogged in-line filter, The number after the error means (0)=Primary pump, (1)=Accumulator pump.
FIX or WORKAROUND
Replace In-Line Filter: part number WAT088084 (also available as pack of 5 as WAT005139)
ADDITIONAL INFORMATION
With the vent valve open, solvent exits the system prior to reaching the filter. If the system initializes with the vent valve open, a clogged in-line filter is the most likely cause.
id2816, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2790XE, 2790XEUP, 2795E, 2795UP, 2795XE, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB