Article number: 279771

SYMPTOMS

  • Squeaking noise comes from the degasser
  • No errors

ENVIRONMENT

  • HPLC
  • UPLC

CAUSE

The degasser exhaust tubing is kinked, restricting the exhaust.

FIX or WORKAROUND

Straighten the tan degasser exhaust tubing to remove any kinks, and ensure that the tube end is not blocked or submerged.

ADDITIONAL INFORMATION