ACQUITY SM-FTN shows random errors during initialization - WKB27915

Article number: 27915

SYMPTOMS

  • Z axis move h/w fault
  • Z axis h/w lost steps
  • Needle arm move h/w fault
  • Tray move fault due to foot (zp) not home
  • Column active preheater over-temp fuse h/w fault)

 

ENVIRONMENT

  • ACQUITY SM-FTN 

CAUSE

Faulty needle carriage    

FIX or WORKAROUND

Replace the needle carriage (700005225)