Skip to main content
Waters

H-Class fails SQT test injections; retention times shifted 0.5 minutes faster and peak height was 0.1 - WKB27913

Article number: 27913

SYMPTOMS

  • After PM, H-Class fails SQT test injections with shifted RT and peak height of 0.1 AU
  • Repeating PM does not correct the issue
  • Adjusting optics and replacing flow cell on TUV does not correct issue
  • Use of premixed mobile phase passes tests
  • Replacing GPV does not correct the issue
  • All diagnostics pass

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC Tunable UV Detector
  • SystemsQT

CAUSE

Wrong solvent (acetone instead of acetonitrile) was used in mobile phase.

FIX or WORKAROUND

  1. Remake mobile phase according to procedure.
  2. Rerun all testing.

ADDITIONAL INFORMATION

 

id27913, eluent, EMP2QLIC, EMP3Q1ULIC, EMP3QLIC, gradient proportioning valve, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

 

Not able to find a solution? Click here to request help.