Accumulator pump head is unable to maintain pressure during dynamic leak test - WKB277841

Article number: 277841

SYMPTOMS

  • Dynamic leak test failure
  • Accumulator pump unable to maintain pressure 
  • Seal wash primed, pump primed; pumps pressurized and delta psi is low 
  • Fittings and check valves were tightened, and no visible signs of leaks; accumulator still unable to maintain pressure 

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

CAUSE

Faulty accumulator check valve

FIX or WORKAROUND

  1. Replace the accumulator check valve
  2. Prime and pressurize the pump head. 
  3. Rerun the dynamic leak test. 