M-Class driver pack fails to install; error code = 1155 - WKB27315

Article number: 27315

SYMPTOMS

  • Waters Deployment Manager fails to install drivers for the µBinary Solvent Manager and µSample Manager:

image.png

ENVIRONMENT

  • ACQUITY UPLC M-Class
  • MassLynx
  • ACQUITY UPLC M-Class Driver Pack v1.53

CAUSE

Extracting the driver pack on a PC with German and English languages corrupts the "µ" in the MSI driver file name. The file names should be:

  • µBinary Solvent Manager
  • µSample Manager - FL

But they are extracted as:

  • ╡Binary Solvent Manager
  • ╡Sample Manager - FL

FIX or WORKAROUND

Extract the driver pack on a different PC with only English language installed and copy the extracted folder to the MassLynx PC.