Alliance 2695 shows poor reproducibility on small-volume injections with 250-µL syringe and 200-µL sample loop - WKB2701
Article number: 2701
SYMPTOMS
- High %RSD failure while making 5-µL injections on an Alliance 2695 with standard 250-µL syringe and 200-µL sample loop
- One of the area counts in a series of six 5-µL injections is much lower
- System precision passes when making injections at 50 µL, 100 µL, and 200 µL
- Seal pack geometry calibration is successful
- Purge and compression test passes
ENVIRONMENT
- Alliance 2695
CAUSE
- Bad sample loop installed at PM. Suspect poorly connected fittings or a sample loop volume that is much larger than 200 µL.
- Sample loop not configured properly
FIX or WORKAROUND
- Replace the 200-µL sample loop (430001630; TUBE ASSY, SAMPLE LOOP, 200 µL, 2695 [for use with phase II injector assembly])
- Verify the sample loop is configured to the correct volume
ADDITIONAL INFORMATION
The Alliance 2695 had a "Phase II Injector". The cut-over to the phase II injector design began in October 2005, beginning with the following serial numbers:
- Part Number / Serial Number / Type
- WAT270008 / J05SM4035M / 2695 w/ sample htr/clr
- WAT270886 / J05SM7373M / 2695 w/o sample htr/clr
- Alternative part number WAT045584 for the "717 SAMPLE LOOP", which is the 200-µL sample loop used on the Alliance 2695 with the original injector assembly
