Skip to main content
Waters

Poor reproducibility with ACQUITY Arc Sample Manager FTN-R - WKB269054

Article number: 269054

SYMPTOMS

  • High variability on samples
  • Poor %RSD on running the same sample (9%)

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY Premier
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Use of non-preslit septa caps on the vials, which can cause multiple issues (coring/cavitation)

FIX or WORKAROUND

Use the Waters-recommended preslit vial caps.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.