Column pre-heater configured but not detected - WKB268235

Article number: 268235

SYMPTOMS

  • Column pre-heater configured but not detected when setting up or running a method
  • Active pre-heater is disabled in the ACQUITY Console configuration 

ENVIRONMENT

  • ACQUITY UPLC Column Manager- Active
  • ACQUITY UPLC Column Heater- Active 

CAUSE

The Instrument method configuration does not match the physical Instrument configuration. 

FIX or WORKAROUND

  1. In the Instrument method control, select Active Preheater as "Use Console Configuration".

ADDITIONAL INFORMATION

 

id268235, A-30HCM, A-30MCM, APH, UPCHA, UPCHAIVD, UPCMA, UPCMAIVD, UPPCHA, UPPCMA