The 2998 Photodiode Array Detector will monitor the baseline but I see no signal after starting the injection in Empower 3 - WKB26304
Article number: 26304
SYMPTOMS
- The 2998 Photodiode Array Detector will monitor the baseline but I see no signal after starting the injection in Empower 3
ENVIRONMENT
- 2998 Photodiode Array Detector
- Empower 3
CAUSE
Faulty inject start cable
FIX or WORKAROUND
- Replace the inject start cable which connects the 2695 Separations Module and the 2998 Photodiode Array Detector
ADDITIONAL INFORMATION
Prior to replacing the cable, try simply reseating the connections and make sure the wire is not pinched or crimped.
id26304, 2998, 2998B, 2998SFC, A-10PDA, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP