Broken outlet fitting on BSM transducer - WKB262365

Article number: 262365

SYMPTOMS

  • Pressure varies
  • Dynamic leak test failure

ENVIRONMENT

  • ACQUITY UPLC
  • I-Class BSM
  • H-Class QSM
  • H-Class Bio QSM
  • ACQUITY Premier BSM
  • nanoACQUITY
  • M-Class
  • ACQUITY ISM
  • UPC2

CAUSE

Broken outlet fitting on BSM transducer

FIX or WORKAROUND

1. Replace the damaged transducer.

  • 700006045 - Transducer, 20K PSI for I-Class systems
  • 700002594 - Pressure Transducer Assembly, 15K psi (ACQUITY UPLC, H-Class, H-Class Bio, nanoACQUITY, UPC2 (B side), ACQUITY Premier)
  • 700006045 - Transducer Assembly ISM, M-Class
  • 700009006 - Transducer Assembly, Pump A, 15K psi, Parylene-Coated, UPC2 (A side)

2. Replace the damaged tubing.

Damaged outlet tubing for pump B accumulator for Aux part number 700009849

ADDITIONAL INFORMATION