Broken outlet fitting on BSM transducer - WKB262365
Article number: 262365
SYMPTOMS
- Pressure varies
- Dynamic leak test failure
ENVIRONMENT
- ACQUITY UPLC
- I-Class BSM
- H-Class QSM
- H-Class Bio QSM
- ACQUITY Premier BSM
- nanoACQUITY
- M-Class
- ACQUITY ISM
- UPC2
CAUSE
Broken outlet fitting on BSM transducer
FIX or WORKAROUND
1. Replace the damaged transducer.
- 700006045 - Transducer, 20K PSI for I-Class systems
- 700002594 - Pressure Transducer Assembly, 15K psi (ACQUITY UPLC, H-Class, H-Class Bio, nanoACQUITY, UPC2 (B side), ACQUITY Premier)
- 700006045 - Transducer Assembly ISM, M-Class
- 700009006 - Transducer Assembly, Pump A, 15K psi, Parylene-Coated, UPC2 (A side)
2. Replace the damaged tubing.
ADDITIONAL INFORMATION
id262365, DLT, UPAISM, UPBSMIVD, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM+, UPIBSMIVD, UPISM, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPLCCOLHTR, UPMASM, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNBINARY, UPNSAMMGR, UPNSOLMGR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR