Article number: 260967

SYMPTOMS

  • ACQUITY TUV does not meet the acceptance criteria for drift
  • The noise specification is met 
  • Lamp energy is good
  • High-quality solvents used

ENVIRONMENT

  • ACQUITY UPLC Tunable UV Detector
  • SystemsQT

CAUSE

Leaking or faulty flow cell

FIX or WORKAROUND

Replace the flow cell.

ADDITIONAL INFORMATION

205015016 - ACQUITY UPLC ID Cell TUV, Analytical.

800000584 - Refurb ACQ TUV Flowcell, Analytical 10 mm (US and CA ONLY)

id260967, A-30UV, EMP2QLIC, EMP3Q1ULIC, EMP3QLIC, UPARPTUV, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC