ACQUITY TUV fails drift test - WKB260967
Article number: 260967
SYMPTOMS
- ACQUITY TUV does not meet the acceptance criteria for drift
- The noise specification is met
- Lamp energy is good
- High-quality solvents used
ENVIRONMENT
- ACQUITY UPLC Tunable UV Detector
- SystemsQT
CAUSE
Leaking or faulty flow cell
FIX or WORKAROUND
Replace the flow cell.
ADDITIONAL INFORMATION
205015016 - ACQUITY UPLC ID Cell TUV, Analytical.
800000584 - Refurb ACQ TUV Flowcell, Analytical 10 mm (US and CA ONLY)