ACQUITY UPLC BSM is dimmed (unavailable) in Empower Run Samples and Console - WKB259075
Article number: 259075
SYMPTOMS
- BSM is dimmed (unavailable) in Run Samples
- BSM appears in Console, but the only function available is "Reset BSM" under "Control"
- "Reset BSM" has no effect
- Console Log shows multiple "Communication Failure" for the BSM
- Node Properties shows "Yes" in the "OK" column for the BSM
ENVIRONMENT
- ACQUITY UPLC BSM
- ACQUITY UPLC I-Class BSM
CAUSE
The BSM ICS has become corrupted.
FIX or WORKAROUND
- Uninstall the BSM ICS software.
- Reboot the PC or LAC/E.
- Reinstall the BSM ICS software.
- Reboot the PC or LAC/E.
ADDITIONAL INFORMATION
id259075, comms, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM