ACQUITY UPLC BSM is dimmed (unavailable) in Empower Run Samples and Console - WKB259075

Article number: 259075

SYMPTOMS

  • BSM is dimmed (unavailable) in Run Samples
  • BSM appears in Console, but the only function available is "Reset BSM" under "Control"
  • "Reset BSM" has no effect
  • Console Log shows multiple "Communication Failure" for the BSM
  • Node Properties shows "Yes" in the "OK" column for the BSM

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC I-Class BSM

CAUSE

The BSM ICS has become corrupted.

FIX or WORKAROUND

  1. Uninstall the BSM ICS software.
  2. Reboot the PC or LAC/E.
  3. Reinstall the BSM ICS software.
  4. Reboot the PC or LAC/E.

ADDITIONAL INFORMATION

 

id259075, comms, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM