M-Class SM-FL error messages during injection: drawing sample rate excessive, sample pressure low, and sample pressure high - WKB256335
Article number: 256335
SYMPTOMS
- Alternating error messages during sample injection: "Warning: Drawing sample rate excessive (-14.1)", "Fatal: Sample pressure low (-10.3): possibly no wash solvent or needle damaged"
- Error after reset and re-initialization: "Sample fluidics high pressure limit (226.4)"
ENVIRONMENT
- M-Class SM-FL
CAUSE
Blocked loop
FIX or WORKAROUND
- Exchange the loop.
- Run the characterize needle seal and characterize needle and loop volume diagnostics as well as the syringe leak tests.
ADDITIONAL INFORMATION