With leak rate rising, primary pump fails dynamic leak test - WKB255351
Article number: 255351
SYMPTOMS
- With leak rate rising as shown, primary pump fails dynamic leak test:
ENVIRONMENT
- ACQUITY Binary Solvent Manager
- ACQUITY Quaternary Solvent Manager
CAUSE
Faulty accumulator check valve
FIX or WORKAROUND
Replace the accumulator check valve.