Peaks are splitting, shouldering, or tailing- WKB25477

Article number: 25477

SYMPTOMS

  • All of the peaks in the chromatogram are split
  • There are no void volumes
  • There are no leaks

ENVIRONMENT

  • 2695 Separations Module
  • 2998 Photodiode Array Detector
  • ACQUITY UPLC
  • I-class
  • Xevo TQD
  • Xevo TQ-S Micro

CAUSE

The column or guard column is faulty.

FIX or WORKAROUND

  1. Replace the guard column.
  2. Replace the column.

ADDITIONAL INFORMATION