Needle/seal move limit exceeded (5.0) - WKB253082
Article number: 253082
SYMPTOMS
- "Needle/seal move limit exceeded" error
- "Seal force sensor h/w fault" error
- No issues with the spring under the seat port
ENVIRONMENT
- ACQUITY I-Class SM-FTN
- ACQUITY H-Class SM-FTN
CAUSE
Needle not fitted correctly on needle carriage

This is what the sample needle should look like when installed correctly.

FIX or WORKAROUND
Reinstall the needle properly and rerun calibrations