Accumulator over pressure on Alliance - WKB248472
Article number: 248472
SYMPTOMS
- Accumulator over pressure and system over pressure errors when at low flow
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
CAUSE
Blockage in the injector
FIX or WORKAROUND
Rebuild seal pack:
Phase II Injectors - WAT271019
Phase III injectors - 700011783
ADDITIONAL INFORMATION
id248472, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB