Accumulator over pressure on Alliance - WKB248472

Article number: 248472

SYMPTOMS

  • Accumulator over pressure and system over pressure errors when at low flow

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module

CAUSE

Blockage in the injector

FIX or WORKAROUND

Rebuild seal pack:

Phase II Injectors - WAT271019

Phase III injectors - 700011783