QSM-R gives continuous beep at power-up - WKB248206
Article number: 248206
SYMPTOMS
- QSM-R gives continuous beep at power-up
- There is no any error message in the log
- The module can work functionally with beeps
ENVIRONMENT
- Arc QSM-R/Bio QSM-R
- H-Class QSM/Bio QSM
- APC pISM/pQSM
- Premier QSM
- Arc Premier QSM
- Arc HPLC QSM-R
CAUSE
Leak sensor circuit
FIX or WORKAROUND
- Turn off the pump.
- Unplug the leak sensor.
- Turn the pump on.
- After initialization, plug the leak sensor back in.
- Set the customer preferences.
- Confirm function.
- Return to customer.
ADDITIONAL INFORMATION
If the procedure above does not resolve the issue, replace the personality board PCB (700005156).