UPLC high pressure - WKB246270

Article number: 246270

SYMPTOMS

  • System high pressure increases until overpressure occurs
  • High pressure remains when disconnecting column
  • Pressure decreases when disconnecting APH from injection valve

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class
  • ACQUITY PREMIER

CAUSE

Clog in APH tubing

FIX or WORKAROUND

  1. Backflush to try to remove the obstruction.
  2. Replace APH. 
    • For H-Class, use part 205000730 (ACQUITY UPLC APH, SS, 12.5LG)
    • For I-Class, use part 205000894 (ACQ. I-Class APH, SS, 12.5 LG, .003 ID)
    • For ACQUITY PREMIER, use 205002234 (ACQUITY APH, MP35N, 12.5" HPS)

ADDITIONAL INFORMATION