Noise does not meet specified value during Noise and Drift test on TUV; drift does meet the value; a dark current error occurs upon power cycle - WKB245679
Article number: 245679
SYMPTOMS
- Noise does not meet the specified value during the Noise and Drift test on TUV; drift does meet the value; a dark current error occurs when power is cycled
- Resetting the dark current error brings successful initialization, and analysis can be done, but the noise is still too high and the Noise and Drift test fails
- All other OQ tests pass
- Replacing the lamp or the cell does not change the situation
ENVIRONMENT
- Acquity TUV
- 2489
CAUSE
Defective part (700009350 - ASSY, PRE-AMP WITH CABLE 2489)
FIX or WORKAROUND
Replace part 700009350 (ASSY, PRE-AMP WITH CABLE 2489).