"Transfer Shuttle Homing Error" on ACQUITY UPLC Sample Organizer - WKB24447

Article number: 24447

SYMPTOMS

  • When taking a plate from a shelf or the XTS onto the y-axis transfer shuttle, the plate/vial holder becomes jammed against the shelf
  • Issue seen only on certain shelves, not all
  • Shelf stabilization kit is fitted

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer

CAUSE

Worn rail on the right-hand side of the shelf

FIX or WORKAROUND

  1. Swap the shelf for another in the organizer.
  2. Replace the shelf with a new one (700009688).

ADDITIONAL INFORMATION