Error with ACQUITY Arc QSM-R: Degasser H/W fault - Vacuum at 1.7PSI(a) - WKB241019

Article number: 241019

SYMPTOMS

  • "Degasser H/W fault - Vacuum at 1.7PSI(a)"

ENVIRONMENT

  • ACQUITY Arc QSM-R

CAUSE

The beige degasser waste line is pinched in the column heater door.

FIX or WORKAROUND

Un-pinch the waste line.

ADDITIONAL INFORMATION

 