Error with ACQUITY Arc QSM-R: Degasser H/W fault - Vacuum at 1.7PSI(a) - WKB241019
Article number: 241019
SYMPTOMS
- "Degasser H/W fault - Vacuum at 1.7PSI(a)"
ENVIRONMENT
- ACQUITY Arc QSM-R
CAUSE
The beige degasser waste line is pinched in the column heater door.
FIX or WORKAROUND
Un-pinch the waste line.