How to disassemble an ELS detector - WKB240843

Article number: 240843

OBJECTIVE or GOAL

Shut down and disassemble an ELS detector for shipment to the repair center.

ENVIRONMENT

  • 2420 Evaporative Light Scattering Detector
  • 2424 Evaporative Light Scattering Detector
  • ACQUITY UPLC ELS Detector

PROCEDURE

  1. Wash the system with H2O for 20 minutes, and then wash the system with IPA for 20 minutes.
  2. Close the control software and power-off the module.
    Electrically unplug the module and mark and remove any cables (Ethernet, signal, I/O, events) from the module.
    (Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of cable and tubing are also helpful.)

                                      2420 Evaporative Light Scattering Detector                                                            2424 Evaporative Light Scattering Detectorimageimage

                                                     ACQUITY UPLC ELS Detector
     image
  3. Disconnect inlet and outlet tubing from the flow cell.
  4. Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
    Note the dimensions and weight of the module to choose an appropriate box:    

    2420 Evaporative Light Scattering Detector: 20.8 cm x 28.4 cm x 50.8 cm, 1.6 kg
    2424 Evaporative Light Scattering Detector: 20.8 cm x 34.3 cm x 61.0 cm, 12.6 kg
    ACQUITY UPLC ELS Detector 21.6 cm x 34.3 cm x 51.8 cm, 14.7 kg  
  5. Fill and sign the latest revision of the H&S module provided by Waters.
  6. Attach it to the box and send a PDF copy to the local Waters office.

  7. The Waters office will contact you to arrange the shipment.                                                                                                                                                          

ADDITIONAL INFORMATION

To reinstall the module, attach all of the cables and tubes that were removed.

Not able to find a solution? Click here to request help.