UPC2 Solvent Manager fails leak test on primary head - WKB239693
Article number: 239693
SYMPTOMS
-
UPC2 solvent pump (B) fails leak test
ENVIRONMENT
- ACQUITY UPC2
CAUSE
Faulty primary check valve
FIX or WORKAROUND
Replace the CO2 head primary check valve - (ASSY, CARTRIDGE, INTELLIGENT VLV, (2PK)) part number - 700003563