What are the default compression check parameters for an Alliance 2695? - WKB239480

Article number: 239480

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

ANSWER

The default parameters listed below are the proper values when performing a compression test for a 2695 Separations Module.

All instruments should pass this test using these parameters.

    Compression Pressure: 300 psi
    Compression Volume: 20.00 µL
    Decay Time: 1.00 minute
    Max Delta Decay: 50.0 psi