Capillary between column and MS diverter valve pops out - WKB239455
Article number: 239455
SYMPTOMS
- Capillary from column pops out of MS diverter valve
ENVIRONMENT
- Xevo TQ-S micro
- Xevo TQD
CAUSE
Clogged or partially clogged ESI probe or fluidics capillaries
FIX or WORKAROUND
- Replace fluidics tubings (700005820), capillaries going in and out of the MS diverter valve.
- Run flow to see if any of the capillaries comes loose.
- If column-to-diverter valve pops out, rebuild the ESI probe (700004672).
ADDITIONAL INFORMATION
id239455, Electrospray, UPCMIVD, UPCOLMGR, XEVOTQD, XEVOTQDIVD, XEVOTQSMIC