Article number: 237947

OBJECTIVE or GOAL

Shut down and disassemble a HPLC RI Detector for shipment to the repair center.

ENVIRONMENT

  • 2410 Differential Refractometer Detector
  • 2414 Differential Refractometer Detector

PROCEDURE

  1. Wash the system in Purge mode with H2O for 20 minutes, and then wash the system in Purge mode with IPA for 20 minutes.
  2. Close the control software and power-off the module.
  3. Electrically unplug the module and mark and remove any cables (Ethernet, signal, I/O, events,IEEE Interface connection) from the module rear.
    (Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of cable and tubing are also helpful.)

clipboard_e5eb053328d1430803a8da5903315e8cc.png         clipboard_ec0679a2305bfb8892ba410912b2e9409.png

4. Disconnect inlet and outlet tubing from the module front and slide off valve outlet tubing from waste line.

clipboard_e4fd219c421fc7a73965414769827469f.png    clipboard_e14a771997f1fc933fbedbe0dcf1017d8.png

5. Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
  Note the dimensions and weight of the module to choose the appropriate box: 

HPLC 2410 Detector 28.4 cm X 20.8 cm X 50.3 cm, 9.7 kg 

HPLC 2414 Detector 28.4 cm X 20.8 cm X 50.3 cm, 11.4 kg

(part number 290001213)

6. Fill and sign the latest revision of the H&S module provided by Waters.

7. Attach it to the box and send a PDF copy to the local Waters office.

8. The Waters office will contact you to arrange the shipment.

ADDITIONAL INFORMATION

To reinstall the module, attach all of the cables and tubes that were removed.

Not able to find a solution? Click here to request help.