Leak sensor does not detect leak from a column fitting on ACQUITY UPLC H-Class Sample Manager FTN system - WKB236991
Article number: 236991
SYMPTOMS
- Visual leak observed from the fitting at the head of the column
- Leak sensor well in the drip tray of the ACQUITY UPLC H-Class Sample Manager FTN does not trigger a Leak Detected error
- Booting the Sample Manager FTN does not resolve the issue
- ACQUITY UPLC H-Class Quaternary Solvent Manager leak sensor also found to not trigger a Leak Detected error
ENVIRONMENT
- ACQUITY UPLC H-Class Sample Manager FTN
- ACQUITY UPLC H-Class Quaternary Solvent Manager
- MassLynx
- ACQUITY MS Console
CAUSE
Leak sensors are disabled in the ACQUITY MS Console.
FIX or WORKAROUND
- Open the ACQUITY MS Console.
- Highlight the system in the system list on the left-hand side.
- Under the menus, go to Control > Leak Sensor.
- Enable the leak sensors.
ADDITIONAL INFORMATION
If you cannot enable all of the leak sensors, close the console and MassLynx. Reboot the workstation and power-cycle the ACQUITY UPLC components. Repeat the steps above.