How to park the vent valve on an ACQUITY UPLC solvent manager - WKB23585

Article number: 23585

OBJECTIVE or GOAL

Park the vent valve on an ACQUITY UPLC solvent manager to align the cartridge with the drive assembly and allow cartridge replacement.

ENVIRONMENT

  • ACQUITY UPLC

PROCEDURE

  1. In the console, go to the interactive display page for the appropriate solvent manager.
  2. On the right-hand side, click to allow control of vent valve positioning.
  3. Select the Vent option under the vent valve position.