Multiple 2695 separations modules are used in the same way, but only one in particular is prone to pressure fluctuations - WKB235395
Article number: 235395
SYMPTOMS
- Multiple 2695 Separations Modules are used in the same way, but only one in particular is prone to pressure fluctuations
ENVIRONMENT
- 2695 Separations Module
CAUSE
1. Defective check valve
2. Defective solvent filter
3. The mobile phase in the unused lines has a little contact with the mobile phase at the GPV.
4. Contamination of dust or the like in the mobile phase bottle
FIX or WORKAROUND
- Clean or replace the check valves.
- Replace the solvent filter (the filter at the end of the tube in the solvent bottle).
- If lines other than the mobile phase line are 100% acetonitrile, change to methanol/water, etc.
- Cap the mobile phase bottle.