UPLC Leak sensor is not recognized after cleaning the sensor following a leak - WKB234612
Article number: 234612
SYMPTOMS
- Leak sensor in the Sample Manager
- Leak sensor not recognized after the leak is cleaned up
ENVIRONMENT
- ACQUITY UPLC I-Class Sample Manager - FTN
- ACQUITY UPLC I-Class Sample Manager - FL
CAUSE
Ribbon cable on the leak sensor is broken.
FIX or WORKAROUND
Replace the leak detector (205000505)
ADDITIONAL INFORMATION
id234612, A-30SM, UPBSMIVD, UPIBSMIVD, UPISMFTN, UPLCCOLHTR, UPNBINARY, UPPFTN, UPSMFTN, UPSMFTN+, UPSMIVD, UPSMPMGR