System overpressures at relatively low psi - WKB233405

Article number: 233405

SYMPTOMS

  • System overpressures at 300 psi and above

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module

CAUSE

The currently stored method's high pressure limit is lower than the normal expected psi.

FIX or WORKAROUND

Do one of the following"

  • Via the front panel of the Alliance, choose default as the method.
  • Edit the instrument method via the front panel of the Alliance to increase the maximum pressure.
  • Power cycle the Alliance to get the system to default to normal pressures.
  • In Empower or MassLynx, click setup new method to get the flow going. This sets the maximum pressure to what is in the method.

ADDITIONAL INFORMATION

 