The T7 tubing into the needle wash station damaged - WKB233030
Article number: 233030
SYMPTOMS
- The T7 needle wash tubing into the needle wash station damaged causing a leak into the sample compartment when priming purging the sample manager.
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
CAUSE
Tubing damaged causing a leak into sample compartment
FIX or WORKAROUND
- Trim the tubing and replace the fittings.
- If the tubing is not long enough, replace it with part 430002346 (ASSY, TUBE, FEED, INJECTION PORT).