Skip to main content
Waters

The T7 tubing into the needle wash station damaged - WKB233030

Article number: 233030

SYMPTOMS

  • The T7 needle wash tubing into the needle wash station damaged causing a leak into the sample compartment when priming purging the sample manager.

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

CAUSE

Tubing damaged causing a leak into sample compartment

FIX or WORKAROUND

  1. Trim the tubing and replace the fittings.
  2. If the tubing is not long enough, replace it with part 430002346 (ASSY, TUBE, FEED, INJECTION PORT).

ADDITIONAL INFORMATION

clipboard_e22c3e52c665d5644241607d79e73668c.png

id233030, UPAQSM, UPARBQSM, UPARBSMFTN, UPARBSMFTN, UPARCH30, UPARHQSM, UPARHSMFTN, UPARPFTN, UPARQSM, UPARSMFTN, UPARSMFTN, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPDAARCB, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.