Skip to main content
Waters

Shifting retention time and poor repeatability with ACQUITY - WKB232745

Article number: 232745

SYMPTOMS

  • Retention times are not stable
  • Repeat injections are not reproducible

ENVIRONMENT

  • ACQUITY UPLC

CAUSE

Retention time issue is due to a leaking binary solvent pump.

Poor peak reproducibility is due to a leaking injector pod on the sample manager.

FIX or WORKAROUND

  1. Replace the seals or check valves as required to get the pump leak-tight up to the vent valve.
  2. Run a dynamic leak test up to the column and replace any leaking tubing or valves as required to pass the leak test.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.