Shifting retention time and poor repeatability with ACQUITY - WKB232745
Article number: 232745
SYMPTOMS
- Retention times are not stable
- Repeat injections are not reproducible
ENVIRONMENT
- ACQUITY UPLC
CAUSE
Retention time issue is due to a leaking binary solvent pump.
Poor peak reproducibility is due to a leaking injector pod on the sample manager.
FIX or WORKAROUND
- Replace the seals or check valves as required to get the pump leak-tight up to the vent valve.
- Run a dynamic leak test up to the column and replace any leaking tubing or valves as required to pass the leak test.
ADDITIONAL INFORMATION
id232745, checkvalve, DLT, UPBINARY