What is the gas flow rate on a 2424 ELSD detector? - WKB23120

Article number: 23120

ENVIRONMENT

  • 2424 Evaporative Light Scattering Detector

ANSWER

With high flow rate (Standard) nebulizer (1.80 standard liters/minute at 25 psi), you can run 300 to 3000 µL/min.

With low flow rate nebulizer (0.77 liters/minute at 25 psi), you can run 50 to 500 µL/min.

 

Click here for nebuliser part numbers 

ADDITIONAL INFORMATION

 