2432 inlet/outlet port is leaking - WKB230911

Article number: 230911

SYMPTOMS

  • Leak at 2432 inlet/outlet port    
  • A tight fitting cannot be made from the capillary tubing to the port

ENVIRONMENT

  • 2432 Conductivity Detector

CAUSE

The threads of the inlet/outlet port of the 2432 are worn out.

clipboard_e6923ef94253a2b4a664de4f0ded3b001.png

FIX or WORKAROUND

  1. The 2432 is an OEM by Metrohm, and the PEEK inlet/outlet ports are not supplied to Waters as spares. 
  2. Use a Waters union to connect them. Explain to the customer that it will not look the same as before.
  3. Any union should work. If you need a union part number, 415000106 is the same union that comes with the Waters MS probe rebuild spare kit. 

     

ADDITIONAL INFORMATION