Error on FTN: Seal residual force limit exceeded (z-wash) (x.x) - WKB230730
Article number: 230730
SYMPTOMS
- "Seal residual force limit exceeded (z-wash)" error on FTN
- Needle seal characterization gives "Z-axis calibration out of range" error
- The load cell at the bottom of wash station responds to pressure and shows increased seal force value on ACQUITY console when pushing down on wash station
- Needle wash and purge lines have solvent in them
- Needle is not bent
- Intermittent "drawing sample rate excessive" warnings in the console log
ENVIRONMENT
- ACQUITY APC Sample Manager - FTN
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC
- ACQUITY Premier
CAUSE
The needle placement (from bottom) setting in the inlet method is too low (needle bottoming out). This results in loss of z steps, intermittent needle blockage, and other errors.
FIX or WORKAROUND
Adjust needle placement higher, until it no longer bottoms out on vial or plate. The default needle placement (automatic) is 4 mm.
ADDITIONAL INFORMATION