Article number: 230730

SYMPTOMS

  • "Seal residual force limit exceeded (z-wash)" error on FTN
  • Needle seal characterization gives "Z-axis calibration out of range" error
  • The load cell at the bottom of wash station responds to pressure and shows increased seal force value on ACQUITY console when pushing down on wash station
  • Needle wash and purge lines have solvent in them
  • Needle is not bent
  • Intermittent "drawing sample rate excessive" warnings in the console log

ENVIRONMENT

  • ACQUITY APC Sample Manager - FTN
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC
  • ACQUITY Premier

CAUSE

The needle placement (from bottom) setting in the inlet method is too low (needle bottoming out). This results in loss of z steps, intermittent needle blockage, and other errors.

FIX or WORKAROUND

Adjust needle placement higher, until it no longer bottoms out on vial or plate. The default needle placement (automatic) is 4 mm.

ADDITIONAL INFORMATION

Needle Placement.PNG