How to configure ACQUITY with Thermo Fisher Exploris MS 480 - WKB226749
Article number: 226749
OBJECTIVE or GOAL
Configure an ACQUITY M-Class with Exploris 480 MS in Windows 10.
ENVIRONMENT
- ACQUITY UPLC M-Class
- ACQUITY UPLC
- Thermo Fisher Exploris 480 MS
- Thermo LC Device 3.2 SP2
PROCEDURE
- Turn off the LC.
- Confirm LAN settings (IPV6 off, power saving off).
- DHCP Server Config wizard on 172.16.0.101.
- Restart the PC.
- Power-on LC and verify that the M-Class populates correctly into the Waters DHCP Server Configurator (see picture below).
- Check LAN settings > Control Panel > All control Panel Items > Network and Sharing Center > Change adapter settings.
Right-click > Properties > ipv4 > compare settings with the picture below.
- Restart the PC.
- Let the customer restart Orbitrap, and let them confirm communications.
- For LC Device Instrument Configuration, include MS and nanoACQUITY.
- Start Xcalibur and confirm that MS and LC both show "Ready to Download" status.
ADDITIONAL INFORMATION
Exploris is running on 172.16.0.200.
id226749, comms, UPMASM, UPMBSM, UPMSMFL