Broad peaks on Alliance - WKB226336
Article number: 226336
SYMPTOMS
- Peak is approximately twice as broad as normal
- Compression check and/or adjust seals fails
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Wear at seal pack
FIX or WORKAROUND
Replace the seal pack and perform the adjust seals and compression check tests.
ADDITIONAL INFORMATION