ACQUITY UPC2 Sample Manager "Sample Pressure Low: Possibly No Wash Solvent or Needle Damaged" error - WKB224733

Article number: 224733

SYMPTOMS

  • "Sample Pressure Low: Possibly No Wash Solvent or Needle Damaged" when trying to inject
  • Wash syringe leak tests successful
  • Sample syringe leak test successful
  • Needle and loop characterization successful
  • VDD passing at an acceptable rate

ENVIRONMENT

  • ACQUITY UPC2
  • Sample Manager - Fixed Loop

CAUSE

Damage to needle

FIX or WORKAROUND

  1. Replace the needle (700009095). 

ADDITIONAL INFORMATION

Customer replaced Klarez O-ring themselves, but they did not tighten the needle wash port, which caused the needle to bend on initialization. They also dislodged the wash port inlet, causing a small leak, which also would have contributed to the error.