ACQUITY UPLC Sample Organizer does not cool to set point - WKB224549

Article number: 224549

SYMPTOMS

  • ACQUITY UPLC Sample Organizer does not cool to set point

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer
  • MassLynx

CAUSE

The Sample Manager temperature control is set to OFF in the ACQUITY console.

FIX or WORKAROUND

Set the Sample Manager temperature control set point in the ACQUITY console.

ADDITIONAL INFORMATION

 