Split peaks and contamination peaks in blanks - WKB224123
Article number: 224123
SYMPTOMS
- Split/broad peaks on samples
- Peaks on blanks
- Replaced/investigated all possible parts/areas of contamination on LC modules (BSM, SM, CM/CH, column and MS)
ENVIRONMENT
- ACQUITY UPLC
- I-Class
- H-Class
CAUSE
Contaminated mobile phases
FIX or WORKAROUND
- Make up fresh mobile phase using new solvents and additives where possible.
- Ensure that you are not using contaminated source of additive and/or solvent.
ADDITIONAL INFORMATION
To identify solvents as the source of the contamination, use a union between the pump outlet tubing and the column inlet tubing.
Replace one component at a time (solvent, buffers, acid, etc.) to pinpoint the source of contamination.
Contamination of the solvents should appear as high background and a chromatographic peak.