SM-FTN "Tray lost synchronization" error - WKB222449

Article number: 222449

SYMPTOMS

  • "Tray lost synchronization" error occurs on SM-FTN with sample organizer

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class

CAUSE

X-axis.png

When the rotary tray rotates, the first plate or the plate transferred from the sample organizer collides with the X shuttle (red frame), preventing rotation and causing an error.

Vibration during operation loosens the nut (orange mark). The X shuttle position changes and interferes with the plate, preventing movement.

FIX or WORKAROUND

  1. Move the bolts and nuts that determine the stop position of the X shuttle (indicated with orange in the photo above), and adjust the position so that the plate does not interfere when the tray rotates.
  2. After adjusting to a position that does not interfere, tighten the nut